Mitutoyo EV Counter Series 542 for Multi-Gage System - 542-067
Features:
- Up to six gages can be connected to one unit, extendable up to 10 units (60 gages at maximum) using the RS Link function* to facilitate the configuration of a multi-point measurement system. * Refer to "Quick Guide to Precision Measuring Instruments" on page G-32 for details of the RS link.
- A range of output modes to choose from: I/O output for tolerance judgment and segment output, BCD data output and RS-232C output are available.
- Other than normal measurement, peak measurement or differential measurement between gages are available.
Function:
- External Control (Zero-set, Preset etc.)
- Direction switch
- Error display
- Tolerance judgment output
- Diverse data output (RS-232C, BCD, Segment)
- Peak measurement Maximum value, minimum value, runout, and differential measurement between two gages Addition, averaging, maximum value, minimum value, and maximum width
Optional Accessories:
- D-EV External display unit: No.02ADD400
- SPC cable (0.5m): No.02ADD950
- SPC cable (1m): No.936937
- SPC cable (2m): No.965014
- AC adapter: No.02ADN460*
- AC cable (USA): 02ZAA010*
- Terminal connecting cable: No.02ADD930*
* Included in package Order No.
LED display: 8 digits for parameter display (displays settings), 1 for error display
Error message: Overspeed, gage error etc.
External display: Dedicated external display unit D-EV (optional) can be connected.
Function of input switches: Measurement mode switching, parameter setting
Input/ output Tolerance judgment output 1 to 6 channels (L1, L2, L3), open-collector BCD output Parallel BCD output (positive/negative-true logic), open-collector Segment output Function to set on only the terminals corresponding to the counting values, open-collector Control output Normal operation signal (NOM), open-collector Control input Output channel designation (segment, in the BCD mode), presetting, peak value clear, range changeover (at segment output), holding counting value open-collector or no-voltage contact signal (with/without contact point).
Interface:
- RS-232C:
Measurement data output and control input
EIA RS-232C-compatible Use cross cables for home position, DTE (terminal definition).
RS link:
- Max. connecting unit: 10 (6 when using EF counter)
- Connecting cable length: Max. 10m (sum of link cable length)
- Data transfer time: 1sec./60ch (when transmission rate is 19200bps)
Rating:
- Power supply voltage: 12 - 24V DC, terminal block (M3 screw)
- Power consumption: 8.4W or less (700mA max.) Ensure at least 1A is available per unit.
- Operating temperature:(humidity) range 0 to 40°C (RH 20 to 80%, no condensation)
- Storage temperature:(humidity) range –10 to 50°C (RH 20 to 80%, no condensation)
- External dimensions: 144 (W) x 72 (H) x 139 (D) mm
Specifications:
Order No | 542-067 |
Package No. (counter w/ AC adapter | 64PKA139 |
Model | EV-16Z |
Applicable input | Differential square-wave |
Applicable gage | LGF with origin point mark |
No of input channels | 6 |
Resolution | .000005, .00005, .0005" / 0.0005, 0.001, 0.005, 0.01mm |
Mass | 910g |
See all Mitutoyo EV Counters Series 542 for Multi-Gage System
Model EV-16Z for use with: